CASE REPORT
Year : 2015  |  Volume : 3  |  Issue : 1  |  Page : 31-32

A simple technique to retrieve broken guide wire transfixing hip joint


Department of Orthopaedics, Veer Chandra Singh Garhwali Government Institute Medical Science and Research Institute, Srinagar, Pauri Garhwal, Uttarakhand, India

Date of Web Publication28-Apr-2015

Correspondence Address:
Daya Krishna
Department of Orthopaedics, Veer Chandra Singh Garhwali Government Medical Science and Research Institute, Srinagar, Pauri Garhwal - 246 174, Uttarakhand
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2319-2585.155920

Rights and Permissions
  Abstract 

While operating on the hip joint, breaking of guide wire became a challenge for treating orthopedic surgeon. Here we report a simple method of removal, which requires limited resources and does not need sophisticated instrumentation. By using this method, we remove broken guide wire in a patient of sub-trochantric fracture of femur managed by close reduction and internal fixation by proximal femoral nail. During the removal of the guide wire, after placing distal screw in the neck the guide wire comes out with some difficulty in the final view on image intensifier - the guide wire broken and transfixes the hip joint but does not enter the pelvis. After removal of the distal screw, its cannulated drill bit was connected to a T-handle and introduced into the track of distal screw up to the broken distal end of guide wire and slow manual reaming done under C-arm image intensifier to prevent damage to the articular cartilage. This process was repeated three times after cleaning drill bit each time to clear the base of guide wire for grasping its distal end. With the help of Kerryson rounger/grasper, the distal end of the guide wire was grasped and easily removed. The distal screw was placed in its original track and the procedure was completed.

Keywords: Breakage, proximal femoral nail, subtrochantric fracture, threaded guide wire


How to cite this article:
Krishna D, Chand S. A simple technique to retrieve broken guide wire transfixing hip joint. J Orthop Allied Sci 2015;3:31-2

How to cite this URL:
Krishna D, Chand S. A simple technique to retrieve broken guide wire transfixing hip joint. J Orthop Allied Sci [serial online] 2015 [cited 2024 Mar 29];3:31-2. Available from: https://www.joas.in/text.asp?2015/3/1/31/155920


  Introduction Top


Threaded guide wire and cannulated instruments are used for localizing the various fixation devices such as cannulated cancellous screws, dynamic hip screw, recon nails, and proximal femoral nail while operating around hip joint, and the intertrochantric and subtrochantric regions. Retrieval of broken threaded guide wire is difficult and challenging. There are chances of intrapelvic migration, damage to the articular cartilage, and over-reaming during its removal. We removed the broken guide wire with the help of manually controlled drilling at the base of guide wire without any further damage to the articular cartilage with the help of Kerryson rounger/grasper. This method can be used by center with limited resources and lacking of sophisticated instruments[1],[2],[3].


  Case Report Top


A 46-year-old male patient presented to the orthopedic emergency with the history of fall from height and inability to use right lower limb. The patient was diagnosed as close subtrochantric fracture of right femur. An above knee skin traction was applied till operative fixation. Close reduction was performed under spinal anesthesia on fracture table and image intensifier. After making an entry at the greater trochanter, the guide wire was passed into the distal fragment followed by reaming and placement of PFN in the correct position. The threaded guide wire was passed into the neck up to the subchondral area of the head under an image intensifier. The proximal screw was placed after drilling and tapping. After placement of the distal screw, the guide wire was removed by forward unidirectional drill used in the case by pulling it out. On the final view of image intensifier, the broken guide wire transfixing the hip was seen [Figure 1]a. For removal of this broken guide wire, we removed the zig and distal screw in the neck after locking of nail in distal fragment. We connected the cannulated drill bit of distal screw to a T-handle and introduced it through the track of distal screw into the neck and head of femur up to the broken end of the guide wire. Under image intensifier, we manually drilled the base of broken drill bit 3 times, after cleaning drill bit each time, to remove bone debris; it cleared the bone around the base of broken guide wire to grasp it easily. The broken end was grasped easily by the Kerryson rounger/grasper and retrieved successfully [Figure 1]b. This method was successfully used in another case, in which the guide wire broken during drilling for proximal screw.
Figure 1: (a) The broken guide wire transfixing hip, (b) broken piece of the retrieved guide wire

Click here to view



  Discussion Top


Guide wire breakage while operating around hip joint is not a common problem; it is very easy to remove the broken guide wire by our method, which we used successfully in two patients. The possible reasons for breakage of guide wire are as follows: Multiple reuse of guide wire, blockage and jamming of cannulated drill bit and tap, increase in the effective diameter of threaded portion of guide wire and decrease in the effective diameter of lumen of cannulated drill bit due to deposition of bone debris produced during the procedure, bending and distortion of guide wire due to the pushing and pulling force applied during insertion and removal, and unidirectional drilling machine used for both insertion and removal may lead to the breakage, as in our cases. The drill machine move in forward direction the threaded portion of guide wire also moves forward but we pull in the opposite direction to remove the guide wire leading to breakage.

The various method of removal of broken guide wire in literature are as follows: by using 2-mm cannulated drill bit under fluoroscopy; by using dynamic hip screw (DHS) reamer to make entry wide and then removal by kocher forceps or a needle holder; by making a window in femoral neck, and by endoscopic-assisted retrieval method.

By using 2-mm cannulated drill bit, it is not easy to catch the guide wire and there are chances of pelvic migration of the guide wire. Window formation is also not without complication; we cannot use DHS triple reamer in case of proximal femoral nail where the implant is in situ, and endoscopic-assisted method requires sophisticated instrumentation and an expert [4],[5],[6],[7],[8].

In our method, no powered instrument was used with the help of T-handle and cannulated drill bit. We cleared the base of the broken guide wire to make it accessible to catch by using Kerryson rounger/grasper for retrieval under image intensifier with implant in situ. The advantages of our method are that there are less chances of pelvic migration of the guide wire and cartilage damage; because of manually controlled drilling under image intensifier, there is no need to wide entry, and the same track is used with implant in situ.[8]

 
  References Top

1.
Arora S, Maini L, Aggarwal V, Dhal A. Broken guide wire protruding into hip joint: A bone endoscopic-assisted retrieval method. Indian J Orthop 2012:46:109-12.  Back to cited text no. 1
    
2.
Sharma H, Chauhan M, Maini L. A technique to remove a broken guide wire transfixing the hip joint. Acta Orthop Belg 2008;74:683-5.  Back to cited text no. 2
    
3.
Mishra P, Jain P, Aggarwal A, Upadhyay A, Maini L, Gautam VK. Intrapelvic protrusion of guide wire during fixation of fracture neck femur. Injury 2002;33:839-41.  Back to cited text no. 3
    
4.
Mishra P, Gautam VK. Broken guide wire with intrapelvic protrusion: A technique of removal. Injury 2004;35:1324-6.  Back to cited text no. 4
    
5.
Sayegh FE, Tintzas D, Kapetanos GA. Intra pelvic migration of guide pin during fixation of a hip fracture: Who and what is to blame?. Acta Orthop Belg 2003;71:239-41.  Back to cited text no. 5
    
6.
Peivandi MT, Kachooei AR, Nazemian Z. New method to remove a broken guide pin in the hip joint. Orthopedics 2011;34:e685-7.  Back to cited text no. 6
    
7.
Docquier PL, Manche E, Autrique JC, Geulette B. Complications associated with gamma nailing. A review of 439 cases. Acta Orthop Belg 2002;68:251-7.  Back to cited text no. 7
    
8.
Sen RK, Tripathi SK, Aggarwal S, Agarwal A, Goyal T, Tahasildar N, et al. Broken Krischner or guide-wire retrieval: A report of 4 cases. Hip Int 2010;20:551-4.  Back to cited text no. 8
    


    Figures

  [Figure 1]


This article has been cited by
1 Broken Pin Removal From Hip Joint Using Arthroscopic Grasper – A Technical Note and Review of Literature
Shahab Beheshti far, Alireza Moharrami, Seyed Peyman Mirghaderi, SM Javad Mortazavi
Injury. 2022;
[Pubmed] | [DOI]
2 Intraoperative instrument breakage during the orthopedic elective procedures: A retrospective single-center experience
GaneshSingh Dharmshaktu,Navneet Adhikari,Pankaj Mourya,ShailendraSingh Bhandari,Pankaj Singh
Journal of Orthopaedics and Spine. 2020; 8(2): 80
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case Report
Discussion
References
Article Figures

 Article Access Statistics
    Viewed7434    
    Printed501    
    Emailed0    
    PDF Downloaded578    
    Comments [Add]    
    Cited by others 2    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]